Align the ADC
:CALibration:GADC
:CALibration:GADC?
Performs the ADC group of alignments. The query returns a 0 if the
alignments occurred without problems.
The query performs the alignment and returns a zero if the alignment
is successful.
Front Panel
Access:
System, Alignments, Align Subsystem, Align ADC
Align the IF Gain
:CALibration:GAIN:IF
:CALibration:GAIN:IF?
Calculate the curve coefficients for the IF gain DAC.
The query performs the alignment and returns a zero if the alignment
is successful.
Front Panel
Access:
System, Alignments, Align Subsystem, IF
Baseband I/Q Commands
Baseband I/Q - Align the Baseband IQ
:CALibration:GIQ
:CALibration:GIQ?
Performs the IQ group of alignments. The query performs the
alignment and returns a 0 if the alignment is succesful.
Remarks: You must be in the Basic, W-CDMA, cdma2000 mode to
use this command. Use INSTrument:SELect to set the
mode.
History: Added revision A.05.00
Baseband I/Q - IQ Common Mode Response Null
:CALibration:IQ:CMR
:CALibration:IQ:CMR?
Forces a common mode response null on I/Q inputs.