294 Chapter 5
Language Reference
SENSe Subsystem
Adjacent Channel Power—Offset Frequency Power Mode
[:SENSe]:ACP:LIST:POWer
INTeg|PEAK,INTeg|PEAK,INTeg|PEAK,INTeg|PEAK,INTeg|PEAK
[:SENSe]:ACP:LIST:POWer?
Define the power measurement mode for each of the offset frequencies.
You can turn off (not use) specific offsets with the
SENS:ACP:LIST:STATe command.
Factory Preset
and *RST: INTeg, INTeg, INTeg, INTeg, INTeg
Remarks: You must be in the NADC mode to use this command.
Use INSTrument:SELect to set the mode.
Adjacent Channel Power—Offset Frequency Relative Limit
[:SENSe]:ACP:LIST:RLIMit
<rel_powr>,<rel_powr>,<rel_powr>,<rel_powr>,<rel_powr>
[:SENSe]:ACP:LIST:RLIMit?
Set the relative limit on offset frequencies. You can turn off (not use)
specific offsets with the SENS:ACP:LIST:STATe command.
Factory Preset
and *RST:
−45 dB
Range:
−200 to 50 dB
Remarks: You must be in the NADC, cdmaOne, or PDC mode to
use this command. Use INSTrument:SELect to set the
mode.
Adjacent Channel Power—Offset Frequency Control
[:SENSe]:ACP:LIST:STATe OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1,
OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:ACP:LIST:STATe?
Turn measurement on or off for the custom offset frequencies.
Mode Offset A Offset B Offset C Offset D Offset E
NADC
−26 dB −45 dB −45 dB 0 dB 0 dB
PDC
−45 dB −60 dB 0 dB 0 dB 0 dB