Agilent Technologies E4406A VSA Satellite Radio User Manual


 
302 Chapter 5
Language Reference
SENSe Subsystem
Adjacent Channel PowerNumber of Measured Points
[:SENSe]:ACP:OFFSet:LIST:POINts
<integer>,<integer>,<integer>,<integer>,<integer>
[:SENSe]:ACP:OFFSet:LIST:POINts?
Selects the number of data points. The automatic mode chooses the
optimum number of points for the fastest measurement time with
acceptable repeatability. The minimum number of points that could be
used is determined by the sweep time and the sampling rate. You can
increase the length of the measured time record (capture more of the
burst) by increasing the number of points, but the measurement will
take longer. Use [:SENSe]:ACP:POINts to set the number of points
used for measuring the reference channel.
Factory Preset
and *RST:
Range: 64 to 65536
Remarks: The fastest measurement times are obtained when the
number of points measured is 2
n
.
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
Adjacent Channel PowerAutomatic Measurement Points
[:SENSe]:ACP:OFFSet:LIST:POINts:AUTO OFF|ON|0|1,
OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:ACP:OFFSet:LIST:POINts:AUTO?
Automatically selects the number of points for the optimum
measurement speed.
Factory Preset
and *RST:
Remarks: You must be in Basic or cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
Mode Offset A Offset B Offset C Offset D Offset E
Basic &
cdmaOne
1024 1024 1024 1024 1024
Mode Offset A Offset B Offset C Offset D Offset E
Basic &
cdmaOne
ON ON ON ON ON