Chapter 5 389
Language Reference
SYSTem Subsystem
Front Panel
Access:
Preset
Preset Type
Preset
and *RST: Factory - This parameter is persistent, which means
that it retains the setting previously selected, even
through a power cycle.
Remarks: :SYST:PRES:USER:SAVE defines the user preset.
Example: SYST:PRES:TYPE FACT
Front Panel
Access:
System, Pwr On/Preset, Preset Factory User
Set Time
:SYSTem:TIME <hour>,<min>,<sec>
:SYSTem:TIME?
Sets the time of the real-time clock of the instrument.
Hour must be an integer from 0 to 23.
Minute must be an integer from 0 to 59.
Second must be an integer from 0 to 59.
Front Panel
Access:
System, Time/Date, Set Time
Adjust Time
:SYSTem:TIME:ADJust <seconds>
Adjust the instruments internal time by the value entered.
Range: Larger than you should ever need
Example: SYST:TIME:ADJ 3600 will advance the time one hour.
SYST:TIME:ADJ -86400 will back the date up one day,
without changing the time of day (minutes or seconds).
History: In revision A.02.00 and later
Default Unit: seconds