Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 395
Language Reference
TRIGger Subsystem
Remarks: You must be in the Basic, cdmaOne, EDGE (w/GSM),
GSM, iDEN, NADC, PDC, Service mode to use this
command. Use INSTrument:SELect to set the mode.
History: Revision A.03.27 or later
Front Panel
Access:
Mode Setup, Trigger, Frame Timer, Offset
Trigger Holdoff
:TRIGger[:SEQuence]:HOLDoff <time>
:TRIGger[:SEQuence]:HOLDoff?
Set the holdoff time between triggers. After a trigger, another trigger
will not be allowed until the holdoff time expires. This parameter
affects all trigger sources.
Factory Preset
and *RST: 0.0 s
20.0 ms for iDEN
10.0 ms for NADC or PDC
Range: 0.0 to 500.0 ms
Default Unit: seconds
Front Panel
Access:
Mode Setup, Trigger, Trig Holdoff
Video (IF) Trigger Delay
:TRIGger[:SEQuence]:IF:DELay <time>
:TRIGger[:SEQuence]:IF:DELay?
Set the trigger delay when using the IF (video) trigger (after the
Resolution BW filter).
Factory Preset
and *RST: 0.0 s
Range:
500.0 ms to 500.0 ms
100.0 ms to 500.0 ms for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & ARIB)
Default Unit: seconds
Front Panel
Access:
Mode Setup, Trigger, Video (IF Envlp), Delay