Agilent Technologies E4406A VSA Satellite Radio User Manual


 
316 Chapter 5
Language Reference
SENSe Subsystem
cdma2000, W-CDMA (Trial & ARIB) mode
[:SENSe]:ACP:SWEep:TYPE FFT|SWEep
[:SENSe]:ACP:SWEep:TYPE?
Selects the type of sweeping.
Fast (W-CDMA (3GPP) mode only) - the data acquisition is made
with the wide channel integration bandwidth and the time-domain
data is divided into the narrow data to apply FFT. This mode is
faster than the FFT mode but less accurate in power levels.
FFT - the data acquisition is made with the narrow channel
integration bandwidth and apply fast Fourier transform (FFT) to
convert to the frequency doman data.
Sweep - the measurement is made by the swept spectrum method
like the traditional swept frequency spectrum analysis to have better
correlation to the input signal with a high crest factor
(peak/averatge ratio). This mode may take a longer time than the
FFT mode. See [:SENSe]:ACP:SWEep:DETector[:FUNCtion].
Factory Preset
and *RST: FFT
Remarks: You must be in the cdma2000, W-CDMA (3GPP), or
W-CDMA (Trial & Arib) mode to use this command.
Use INSTrument:SELect to set the mode.
Adjacent Channel PowerTrigger Source
[:SENSe]:ACP:TRIGger:SOURce
EXTernal[1]|EXTernal2|FRAMe|IF|IMMediate|RFBurst
[:SENSe]:ACP:TRIGger:SOURce?
Select the trigger source used to control the data acquisitions.
EXTernal 1 front panel external trigger input
EXTernal 2 rear panel external trigger input
FRAMe internal frame trigger from front panel input
IF internal IF envelope (video) trigger
IMMediate the next data acquisition is immediately taken,
capturing the signal asynchronously (also called free run).
RFBurst wideband RF burst envelope trigger that has automatic
level control for periodic burst signals.
Factory Preset
and *RST: IMMediate for BS