Chapter 5 299
Language Reference
SENSe Subsystem
Default Unit: Hz
Remarks: You must be in Basic, cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or iDEN mode to use
this command. Use INSTrument:SELect to set the
mode.
Adjacent Channel Power—FFT Segments
[:SENSe]:ACP:OFFSet:LIST:FFTSegment
<integer>,<integer>,<integer>,<integer>,<integer>
[:SENSe]:ACP:OFFSet:LIST:FFTSegment?
Selects the number of FFT segments used in making the measurement.
In automatic mode the measurement optimizes the number of FFT
segments required for the shortest measurement time. The minimum
number of segments required to make a measurement is set by your
desired measurement bandwidth. Selecting more than the minimum
number of segments will give you more dynamic range for making the
measurement, but the measurement will take longer to execute.
Factory Preset
and *RST:
Range: 1 to 12
Remarks: You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History: Revision A.03.00 or later, in cdmaOne revision A.04.00
Adjacent Channel Power—Automatic FFT Segments
[:SENSe]:ACP:OFFSet:LIST:FFTSegment:AUTO OFF|ON|0|1,
OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1, OFF|ON|0|1
[:SENSe]:ACP:OFFSet:LIST:FFTSegment:AUTO?
The automatic mode selects the optimum number of FFT segments to
make the fastest possible measurement.
Mode Offset A Offset B Offset C Offset D Offset E
Basic &
cdmaOne
1 1111