Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 259
Language Reference
MEASure Group of Commands
Adjacent Channel Power Ratio (ACP) Measurement
This measures the total rms power in the specified channel and in 5
offset channels. You must be in Basic, cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), iDEN, NADC or PDC mode to use
these commands. Use INSTrument:SELect to set the mode.
The general functionality of CONFigure, FETCh, MEASure, and READ
are described at the beginning of this section. See the SENSe:ACP
commands for more measurement related commands.
:CONFigure:ACP
:FETCh:ACP[n]?
:READ:ACP[n]?
:MEASure:ACP[n]?
For Basic mode, a channel frequency and power level can be defined in
the command statement to override the default standard setting. A
comma must precede the power value as a place holder for the
frequency when no frequency is sent.
History: Added to Basic mode, version A.03.00 or later
Front Panel
Access:
Measure, ACP or ACPR
After the measurement is selected, press Restore Meas
Defaults
to restore factory defaults.
Measurement Results Available
Measurement
Type
n Results Returned
0 Returns unprocessed I/Q trace data, as a series of
comma-separated trace points, in volts. The I values are listed
first in each pair, using the 0 through even-indexed values. The
Q values are the odd-indexed values.
not
specified
or n=1
NADC and
PDC mode
Returns 22 comma-separated scalar results, in the following
order:
1. Center frequency absolute power (dBm)
2. Center frequency absolute power (W)
3. Negative offset frequency (1) relative power (dB)
4. Negative offset frequency (1) absolute power (dBm)
5. Positive offset frequency (1) relative power (dB)
6. Positive offset frequency (1) absolute power (dBm)
...
1. Positive offset frequency (5) relative power (dB)
2. Positive offset frequency (5) absolute power (dBm)