Agilent Technologies E4406A VSA Satellite Radio User Manual


 
348 Chapter 5
Language Reference
SENSe Subsystem
For bursty signals, auto-peak lock ranging should not be used.
The measurement will fail to operate, since the wrong (locked)
ADC range will be chosen often and overloads will occur in the
ADC.
M6 - manually selects an ADC range that subtracts 6 dB of fixed
gain across the range. Manual ranging is best for CW signals.
P0 to 24 - manually selects ADC ranges that add 0 to 24 dB of fixed
gain across the range. Manual ranging is best for CW signals.
Factory Preset
and *RST: APEak
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
SpectrumAverage Clear
[:SENSe]:SPECtrum:AVERage:CLEar
The average data is cleared and the average counter is reset.
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
SpectrumNumber of Averages
[:SENSe]:SPECtrum:AVERage:COUNt <integer>
[:SENSe]:SPECtrum:AVERage:COUNt?
Set the number of sweeps that will be averaged. After the specified
number of sweeps (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST: 25
Range: 1 to 10,000
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
SpectrumAveraging State
[:SENSe]:SPECtrum:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:SPECtrum:AVERage[:STATe]?
Turn averaging on or off.