Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 333
Language Reference
SENSe Subsystem
Front Panel
Access:
Input, Max Total Pwr (at UUT)
Power Statistics CCDF Measurement
Commands for querying the statistical power measurement of the
complimentary cumulative distribution function (CCDF) measurement
results and for setting to the default values are found in the MEASure
Group of Commands on page 255. The equivalent front panel keys for
the parameters described in the following commands, are found under
the
Meas Setup key, after the Power Stat CCDF measurement has been
selected from the
MEASURE key menu.
History: Added PSTatistic to Basic Mode version A.04.00
Power Statistics CCDFChannel Bandwidth
[:SENSe]:PSTatistic:BANDwidth|BWIDth <freq>
[:SENSe]:PSTatistic:BANDwidth|BWIDth?
Set the bandwidth that will be used for acquiring the signal.
Factory Preset
and *RST: 5.0 MHz
Range: 10.0 kHz to 6.7 MHz
Resolution: 0.1 kHz
Step: 1.0 kHz
Default Unit: Hz
Remarks: You must be in the Basic, cdma2000, W-CDMA (3GPP),
or W-CDMA (Trial & Arib) mode to use this command.
Use INSTrument:SELect to set the mode.
Power Statistics CCDFSample Counts
[:SENSe]:PSTatistic:COUNts <integer>
[:SENSe]:PSTatistic:COUNts?
Set the counts. Measurement stops when the sample counts reach this
value.
Factory Preset
and *RST: 10,000,000
Range: 1,000 to 2,000,000,000
Unit: counts