CALCulate Subsystem
Marker X Value
:CALCulate:<measurement>:MARKer[1]|2|3|4:X <param>
:CALCulate:<measurement>:MARKer[1]|2|3|4:X?
Position the designated marker on its assigned trace at the specified X
value. The parameter value is in X-axis units (which is often frequency
or time).
The marker must have already been assigned to a trace. Use
:CALCulate:<measurement>:MARKer[1]|2|3|4:TRACe to assign a
marker to a particular trace.
The query returns the current X value of the designated marker. The
measurement must be completed before querying the marker.
Example: CALC:SPEC:MARK2:X 1.2e6 Hz
Default Unit: Matches the units of the trace on which the marker is
positioned
Remarks: The keyword for the current measurement must be
specified in the command. (Some examples include:
SPECtrum, WAVeform)
Front Panel
Access:
Marker, <active marker>, RPG
Marker X Position
:CALCulate:<measurement>:MARKer[1]|2|3|4:X:POSition
<integer>
:CALCulate:<measurement>:MARKer[1]|2|3|4:X:POSition?
Position the designated marker on its assigned trace at the specified X
position. A trace is composed of a variable number of measurement
points. This number changes depending on the current measurement
conditions. The current number of points must be identified before
using this command to place the marker at a specific location.
The marker must have already been assigned to a trace. Use
:CALCulate:<measurement>:MARKer[1]|2|3|4:TRACe to assign a
marker to a particular trace.
The query returns the current X position for the designated marker.
The measurement must be completed before querying the marker.
Example: CALC:SPEC:MARK:X:POS 500
Range: 0 to a maximum of (3 to 920,000)
Remarks: The keyword for the current measurement must be
specified in the command. (Some examples include: