Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 291
Language Reference
SENSe Subsystem
measure the offset channels and wider dynamic range for the
measurement is available.
Factory Preset
and *RST: 0
Range:
40.00 to 0.00 dB
Remarks: You must be in the W-CDMA (3GPP) mode to use this
command. Use INSTrument:SELect to set the mode.
Adjacent Channel PowerRoot Raised Cosine Filter Alpha
[:SENSe]:ACP:FILTer[:RRC]:ALPHa <numeric>
[:SENSe]:ACP:FILTer[:RRC]:ALPHa?
Set the alpha value of the Root Raised Cosine (RRC) filter.
Factory Preset
and *RST: 0.22
Range: 0.01 to 0.5
Remarks: You must be in the W-CDMA (3GPP) mode to use this
command. Use INSTrument:SELect to set the mode.
Adjacent Channel PowerRoot Raised Cosine Filter Control
[:SENSe]:ACP:FILTer[:RRC][:STATe] OFF|ON|0|1
[:SENSe]:ACP:FILTer[:RRC][:STATe]?
Turn the Root Raised Cosine (RRC) filter on or off.
Factory Preset
and *RST: On
Remarks: You must be in the W-CDMA (3GPP) mode to use this
command. Use INSTrument:SELect to set the mode.
Adjacent Channel PowerReference Channel FFT Segments
[:SENSe]:ACP:FFTSegment <integer>
[:SENSe]:ACP:FFTSegment?
Selects the number of FFT segments used in making the measurement
of the reference channel (carrier). In automatic mode the measurement
optimizes the number of FFT segments required for the shortest
measurement time. The minimum number of segments required to
make a measurement is set by your desired measurement bandwidth.