Agilent Technologies E4406A VSA Satellite Radio User Manual


 
324 Chapter 5
Language Reference
SENSe Subsystem
History: Version A.03.00 or later
Front Panel
Access:
FREQUENCY Channel, TSC (Std)
Training Sequence Code (TSC) Auto
[:SENSe]:CHANnel:TSCode:AUTO OFF|ON|0|1
[:SENSe]:CHANnel:TSCode:AUTO?
Select auto or manual control for training sequence code (TSC) search.
With auto on, the measurement is made on the first burst found to have
one of the valid TSCs in the range 0 to 7 (i.e. normal bursts only). With
auto off, the measurement is made on the 1st burst found to have the
selected TSC.
Factory Preset
and *RST: AUTO
Remarks: Global to the current mode.
You must be in the EDGE(w/GSM), GSM mode to use
this command. Use INSTrument:SELect to set the
mode.
Front Panel
Access:
FREQUENCY Channel, TSC (Std)
Channel Power Measurement
Commands for querying the channel power measurement results and
for setting to the default values are found in the MEASure Group of
Commands on page 255. The equivalent front panel keys for the
parameters described in the following commands, are found under the
Meas Setup key, after the Channel Power measurement has been selected
from the
MEASURE key menu. CHPower used instead of the more
std-compliant CPOWer, as that syntax was already used for Carrier
Power measurement (but has since been renamed).
Channel PowerAverage Count
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the averaging mode (terminal
control) setting determines the averaging action.