Agilent Technologies E4406A VSA Satellite Radio User Manual


 
Chapter 5 351
Language Reference
SENSe Subsystem
combinations of these settings that are not valid.
Factory Preset
and *RST: 1.55 MHz
1.25 MHz for cdmaOne
155.0 kHz, for iDEN mode
Range: 1 Hz to 10.0 MHz
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
SpectrumPre-FFT BW Filter Type
[:SENSe]:SPECtrum:BANDwidth|BWIDth:PFFT:TYPE FLAT|GAUSsian
[:SENSe]:SPECtrum:BANDwidth|BWIDth:PFFT:TYPE?
Select the type of pre-FFT filter that is used. This is an advanced
control that normally does not need to be changed.
Flat top (FLAT)- a filter with a flat amplitude response, which
provides the best amplitude accuracy.
GAUSsian - a filter with Gaussian characteristics, which provides
the best pulse response.
Factory Preset
and *RST: FLAT
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
SpectrumResolution BW
[:SENSe]:SPECtrum:BANDwidth|BWIDth[:RESolution] <freq>
[:SENSe]:SPECtrum:BANDwidth|BWIDth[:RESolution]?
Set the resolution bandwidth for the FFT. This is the bandwidth used
for resolving the FFT measurement. It is not the pre-FFT bandwidth.
This value is ignored if the function is auto-coupled.
Frequency span, resolution bandwidth, and the pre-FFT bandwidth
settings are normally coupled. If you are not auto-coupled, there can be
combinations of these settings that are not valid.
Factory Preset
and *RST: 20.0 kHz
250.0 Hz, for iDEN mode
Range: 0.10 Hz to 3.0 MHz