338 Chapter 5
Language Reference
SENSe Subsystem
burst edges.
Factory Preset
and *RST: 1
Range: 1 to 50 (for resolution BW = 500 kHz)
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—Trigger Source
[:SENSe]:PVTime:TRIGger:SOURce EXTernal[1]|EXTernal2
|FRAMe|IF|IMMediate|RFBurst
[:SENSe]:PVTime:TRIGger:SOURce?
Select the trigger source used to control the data acquisitions.
EXTernal 1 - front panel external trigger input
EXTernal 2 - rear panel external trigger input
FRAMe - uses the internal frame timer, which has been
synchronized to the selected burst sync.
IF - internal IF envelope (video) trigger
IMMediate - the next data acquisition is immediately taken,
capturing the signal asynchronously (also called Free Run).
RFBurst - wideband RF burst envelope trigger that has automatic
level control for periodic burst signals.
Factory Preset
and *RST: RFBurst if the RF Burst Hardware (option B7E) has
been installed
EXTernal if option B7E has not been installed
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.