Agilent Technologies E4406A VSA Satellite Radio User Manual


 
362 Chapter 5
Language Reference
SENSe Subsystem
WaveformDecimation of Waveform Display
[:SENSe]:WAVeform:DECimate[:FACTor] <integer>
[:SENSe]:WAVeform:DECimate[:FACTor]?
Set the amount of data decimation done on the IQ data stream. For
example, if 4 is selected, three out of every four data points will be
thrown away. So every 4th data point will be kept.
Factory Preset
and *RST: 1
Range: 1 to 4
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
WaveformControl Decimation of Waveform Display
[:SENSe]:WAVeform:DECimate:STATe OFF|ON|0|1
[:SENSe]:WAVeform:DECimate:STATe?
Set the amount of data decimation done by the hardware in order to
decrease the number of acquired points in a long capture time. This is
the amount of data that the measurement ignores.
Factory Preset
and *RST: OFF
Remarks: To use this command, the appropriate mode should be
selected with INSTrument:SELect.
WaveformSweep (Acquisition) Time
[:SENSe]:WAVeform:SWEep:TIME <time>
[:SENSe]:WAVeform:SWEep:TIME?
Set the measurement acquisition time. It is used to specify the length of
the time capture record.
Factory Preset
and *RST: 2.0 ms
10.0 ms, for NADC, PDC
15.0 ms, for iDEN mode
Range: 1
µs to 100 s
Default Unit: seconds