Agilent Technologies E4406A VSA Satellite Radio User Manual


 
286 Chapter 5
Language Reference
SENSe Subsystem
SENSe Subsystem
Sets the instrument state parameters so that you can measure the
input signal.
The SCPI default for data output format is ASCII. The format can be
changed to binary with FORMat:DATA which transports faster over the
bus.
Adjacent Channel Power Measurement
Commands for querying the adjacent channel power measurement
results and for setting to the default values are found in the MEASure
Group of Commands on page 255. The equivalent front panel keys for
the parameters described in the following commands, are found under
the
Meas Setup key, after the ACP or ACPR measurement has been
selected from the
MEASURE key menu.
Adjacent Channel PowerAverage Count
[:SENSe]:ACP:AVERage:COUNt <integer>
[:SENSe]:ACP:AVERage:COUNt?
Set the number of data acquisitions that will be platform averaged.
After the specified number of average counts, the average mode
(termination control) setting determines the average action.
Factory Preset
and *RST: 10 for cdma2000, W-CDMA (3GPP), W-CDMA (Trial &
Arib)
20 for Basic, cdmaOne, iDEN
Range: 1 to 10,000
Remarks: Use INSTrument:SELect to set the mode.
Adjacent Channel PowerAveraging State
[:SENSe]:ACP:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:ACP:AVERage[:STATe]?
Turn average on or off.
Factory Preset
and *RST: On
Off for iDEN mode
Remarks: Use INSTrument:SELect to set the mode.