336 Chapter 5
Language Reference
SENSe Subsystem
Power vs. Time—Averaging Mode
[:SENSe]:PVTime:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:PVTime:AVERage:TCONtrol?
Select the type of termination control used for the averaging function.
This specifies the averaging action after the specified number of bursts
(average count) is reached.
EXPonential - Each successive data acquisition after the average
count is reached is exponentially weighted and combined with the
existing average.
REPeat - After reaching the average count, the averaging is reset
and a new average is started.
Factory Preset
and *RST: EXPonential
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—Averaging Type
[:SENSe]:PVTime:AVERage:TYPE
LOG|MAXimum|MINimum|MXMinimum|RMS
[:SENSe]:PVTime:AVERage:TYPE?
Select the type of averaging to be performed.
LOG - The log of the power is averaged. (This is also known as video
averaging.)
MAXimum - The maximum values are retained.
MINimum - The minimum values are retained.
MXMinimum - Both the maximum and the minimum values are
retained.
RMS - The power is averaged, providing the rms of the voltage.
Factory Preset
and *RST: RMS
Remarks: You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.