Agilent Technologies E4406A VSA Satellite Radio User Manual


 
190 Chapter 5
Language Reference
Common IEEE Commands
For example, if you want to verify the completion of both calibrating
and waiting for trigger set :STAT:OPER:ENAB 17 and monitor any
changes.
Key Type: There is no equivalent front panel key.
Query Instrument Options
*OPT?
Returns a string of all the installed instrument options.It is a comma
separated list such as: BAC,BAH. There are a few options that include
more then one mode. An istrument with one of these options will report
the option number once for each mode. You would get a response:
BAC,BAE,BAE,BAH For an instrument that contains cdmaOne (BAC),
NADC (BAE), PDC (BAE), and GSM (BAH).
Recall
*RCL <register>
This command recalls the instrument state from the specified
instrument memory register.
Range: registers are an integer, 0 to 19
Front Panel
Access:
File, Recall State
Reset
*RST
This command presets the instrument to a factory defined condition
that is appropriate for remote programming operation. *RST is
equivalent to performing the two commands :SYSTem:PRESet and
*CLS. *RST does not change the mode and only resets the parameters for
the current mode.
Waiting for trigger 5 16
Printing 11 1024
Mass memory access
(floppy drive)
12 2048
VSAProcess STATus:OPER
Register Bit
Byte Value