312 Chapter 5
Language Reference
SENSe Subsystem
Remarks: You must be in Basic, cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or iDEN mode to use
this command. Use INSTrument:SELect to set the
mode.
Adjacent Channel Power—Number of Measured Points
[:SENSe]:ACP:POINts <integer>
[:SENSe]:ACP:POINts?
Selects the number of data points used to measure the reference
(carrier) channel. The automatic mode chooses the optimum number of
points for the fastest measurement time with acceptable repeatability.
The minimum number of points that could be used is determined by the
sweep time and the sampling rate.
You can increase the length of the measured time record (capture more
of the burst) by increasing the number of points, but the measurement
will take longer. Use [:SENSe]:ACP:OFFSet:LIST:POINts to set the
number of points used for measuring the offset channels.
Factory Preset
and *RST: 1024
Remarks: The fastest measurement times are obtained when the
number of points measured is 2
n
.
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
Range: 64 to 65536
Adjacent Channel Power—Automatic Measurement Points
[:SENSe]:ACP:POINts:AUTO OFF|ON|0|1
[:SENSe]:ACP:POINts:AUTO?
Automatically selects the number of points for the optimum
measurement speed.
Factory Preset
and *RST: ON
W-CDMA
(3GPP)
REL REL REL REL REL
W-CDMA (Trial
& Arib)
REL REL REL REL REL
Mode Variant Offset A Offset B Offset C Offset D Offset E