236 Chapter 5
Language Reference
DISPlay Subsystem
n=1, m=1 RF envelope
n=2, m=1 I/Q Waveform
n=2, m=1 I and Q Waveform (Option B7C)
n=4, m=1 I/Q Polar (Basic, W-CDMA, cdma2000)
n=5, m=1 Linear Envelope (Option B7C)
m, selects the window within the view. The default is 1.
Factory Preset: 10 dBm, for RF envelope
Range: .1 dB to 20 dB, for RF envelope
Default Unit: dBm, for RF envelope
Remarks: May affect input attenuator setting.
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Front Panel
Access: When in Waveform measurement:
Amplitude Y Scale,
Scale/Div.
History: Modified revision A.05.00
Waveform - Y-Axis Reference Level
:DISPlay:WAVeform[n]:WINDow[m]:TRACe:Y[:SCALe]:RLEVel
<power>
:DISPlay:WAVeform[n]:WINDow[m]:TRACe:Y[:SCALe]:RLEVel?
Sets the amplitude reference level for the y-axis.
n, selects the view, the default is RF envelope.
n=1, m=1 RF envelope
n=2, m=1 I/Q Waveform
n=2, m=1 I and Q Waveform (Option B7C)
n=4, m=1 I/Q Polar (Basic, W-CDMA, cdma2000)
n=5, m=1 Linear Envelope (Option B7C)
m, selects the window within the view. The default is 1.
Factory Preset: 0 dBm, for RF envelope
Range:
250 to 250 dBm, for RF envelope
Default Unit: dBm, for RF envelope
Remarks: May affect input attenuator setting.