292 Chapter 5
Language Reference
SENSe Subsystem
Selecting more than the minimum number of segments will give you
more dynamic range for making the measurement, but the
measurement will take longer to execute.
To use this command you must first set SENSe:ACP:FFTS:AUTO to off.
Factory Preset
and *RST: 1
Range: 1 to 12
Remarks: You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History: Revision A.03.00 or later, in cdmaOne revision A.04.00
Adjacent Channel Power—Reference Channel FFT Segments
State
[:SENSe]:ACP:FFTSegment:AUTO OFF|ON|0|1
[:SENSe]:ACP:FFTSegment:AUTO?
The automatic mode selects the optimum number of FFT segments to
measure the reference channel (carrier), while making the fastest
possible measurement.
Factory Preset
and *RST: ON
Remarks: You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
History: Revision A.03.00 or later, in cdmaOne revision A.04.00
Adjacent Channel Power—Frequency Span Query
[:SENSe]:ACP:FREQuency:SPAN?
Returns the span of the spectrum view.
Remarks: You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
MEAS|READ|FETC:ACP4? returns the
frequency-domain spectrum trace data for the entire
frequency range being measured..
History: Revision A.05.00 or later