Agilent Technologies E4406A VSA Satellite Radio User Manual


 
382 Chapter 5
Language Reference
SYSTem Subsystem
SYSTem Subsystem
This subsystem is used to set the controls and parameters associated
with the overall system communication. These are functions that are
not related to instrument performance. Examples include functions for
performing general housekeeping and functions related to setting
global configurations.
GPIB Address
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <integer>
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?
Sets and queries the GPIB address.
Example: SYST:COMM:GPIB:ADDR 18
Factory Preset
and *RST: The factory default is 18.
This function is persistent which means that it stays at
the setting previously selected, even through a power
cycle.
Range: Integer, 0 to 30
Example: SYST:COMM:GIPB:ADDRESS 18
Front Panel
Access:
System, Config I/O, GPIB Addr
LAN IP Address with Host Name
:SYSTem:COMMunicate:LAN[:SELF]:IP <string>
:SYSTem:COMMunicate:LAN[:SELF]:IP?
Set the IP (internet protocol) address, domain name and node name for
the instrument.
<string> is a string that contains: <IP address> <host name> as shown
in the following example:
141.4.402.222 sigan
where: 141.4.402.222, is the IP address and sigan, is the host name.
Example: SYST:COMM:LAN:IP 22.121.44.45 analyz
Front Panel
Access:
System, Config I/O, Config LAN