Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 215
Preliminary Basic Digital Operation (Option 651/652/654)
Preliminary Multiple Baseband Generator Synchronization
1. On the master, press the Sync Slaves softkey.
NOTE All of the signal generators in the master/slave system must be resynchronized when any
changes are made to the master/slave settings or with the addition of a slave instrument,
even if In Sync appears after pressing the Listen for Sync softkey on the slave instruments.
2. On the front panel displays, ensure that all of the signal generators show In Sync as the Status.
Trigger and Play the Waveform
1. On all Agilent MXGs, press Mode > Dual ARB > ARB Off On to On.
2. Start the trigger signal going to the master signal generator.
Making Changes to the Multiple Synchronization Setup and Resynchronizing the Master/Slave
System
If any changes are made to the master/slave parameters or a signal generator (slave unit) is added to
the system, the system must be resynchronized even if In Sync appears in the Status portion of the
display.
1. Turn off the trigger source. If using the Trigger key, there is nothing to turn off.
If the trigger source is on and provides a continuous pulse stream, it may cause the signal
generators to incorrectly display In Sync as the status after pressing the Listen for Sync softkey.
2. On each of the signal generators, press Mode > Dual ARB > ARB Off On to Off.
If a signal generator(s) has the Dual ARB on during changes, signal generators further in the
chain may incorrectly display In Sync after pressing the Listen for Sync softkey.
3. On each of the signal generators, press ARB Setup > More > Multi–BBG Sync Setup.
4. Make the changes in the Multi–BBG Sync Setup menu.
Out Of Sync appears as the status message.
5. On each of the slave signal generators, press Listen for Sync.
6. Ensure that all of the slaves’ Status show Waiting for Sync. If In Sync shows as the status,
perform the following steps:
a. Check that the PATT TRIG to EVENT 1 cables are properly connected on the rear panel.
A disconnected cable can cause a false In Sync status.
b. If the cables are connected, perform steps 1 and 2.
c. Press Listen for Sync and ensure that Waiting for Sync appears as the status.
7. On the master signal generator, press Sync Slaves.
8. Verify that In Sync appears as the status on all master/slave signal generators.
9. Perform the process “Trigger and Play the Waveform” on page 215.