Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 45
Preliminary Basic Operation
Preliminary Setting Frequency and Power (Amplitude)
Setting Frequency and Power (Amplitude)
Figure 3-1 Frequency and Amplitude Softkeys
page107
To display the next menu, press More.
For details on each key, use key help
as described on page 42.
page46
Refer to the SCPI Command
Reference.
Opens the
Atten/ALC
Control
menu.
Enables the optimize signal-to-noise ratio state (Optimize S/N) On or Off.
The optimize signal–to–noise softkey changes the attenuator and alc setting to provide optimal
signal–to–noise performance; it does not change the RF output power.
Caution: When the optimize signal-to-noise ratio state is enabled, some increased levels of
harmonic distortion can occur. This increased harmonic distortion could degrade ACPR and EVM.
Note: This mode is mutually exclusive with attenuator hold (Atten Hold), and any modulation type. A
settings conflict error will be generated if attentuator hold or any modulation is activated when
optimize signal–to–noise is active (On).
page 47
This softkey sets a user
determined maximum
power level on the
instrument that is
persistent.
Enables the
User Power
Max softkey.
page88
In Amplitude mode, this menu is
automatically displayed when entering
a numeric value with the front panel
keypad.
In Frequency mode, this menu is
automatically displayed when entering
a numeric value with the front panel
keypad.