Agilent Technologies 62A Stereo System User Manual


 
86 Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide
Optimizing Performance Preliminary
Using the Dual Power Meter Display Preliminary
13. On the MXG: Press Channel B to On and then back to Off again. This initializes the MXG to the
external power meter.
14. Press Return > Zero Sensor
A diagnostic dialog box is displayed each time an external power meter is being used and the
Zero Sensor or Calibrate Sensor softkey is pressed (refer to Figure 5- 10 on page 86).
Verify the power sensor is connected to the 50 MHz reference of the power meter.
Figure 5-10 Diagnostic Dialog Box for Channel B
15. Press Done
The Running Calibration(s) bar is displayed: Zeroing Sensor Please wait....
16. After Running Calibration(s) bar disappears: Press Calibrate Sensor
Diagnostic Dialog box is displayed that prompts for verifying the connection of the power sensor
to the power meter 50 MHz reference (refer to Figure 5- 11 on page 86).
Figure 5-11 Diagnostic Dialog Box for Calibration
For details on
each key, use key
help as described
on see page 42.
For details on
each key, use key
help as described
on see page 42.