210 Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide
Basic Digital Operation (Option 651/652/654) Preliminary
Multiple Baseband Generator Synchronization Preliminary
Figure 8-27 Multiple Baseband Generator Synchronization (BBG Synchronization) Trigger Softkeys and Menu Location
For details on each key, use key help
as described on page 42.
Greys out for slave instruments.
Remains active for the master
instrument.
see page 211
BBG sync automatically changes the trigger type to Trigger and Run
if the current setting is Free Run or Gated. If desired, manually set
differently on either the master or slaves, or both either before or
after configuring the BBG sync setup parameters.
Free Run and Gated are unavailable choices while using BBG sync.
Master default settings. If
desired, change these
settings after designating the
instrument as the master
Slave default settings. These
settings are not accessible
after selecting an instrument
as a slave.
Note: The BBG sync feature automatically configures the trigger settings shown below. To avoid a settings conflict error in this
process, manually configure the trigger settings prior to setting the BBG sync parameters shown on page 211.
see page 134
see page 164
Neg is the factory default setting.
The slave selection changes it to
Pos. If desired, manually change to
Pos prior to selecting as a slave.
Note: A setting conflict error occurs
when at the time of the master/slave
selection, the following trigger setting(s)
is true:
Master
Free Run or Gated trigger type
Slave
Free Run or Gated trigger type
Ext Polarity Neg trigger source