Agilent Technologies 62A Stereo System User Manual


 
Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide 119
Preliminary Optimizing Performance
Preliminary Using a USB Keyboard
Using a USB Keyboard
You can use a USB keyboard to remotely control the RF output state, the modulation state, and to
select a memory sequence and register.
The register selection, RF output state, and modulation state are affected by power cycle or preset,
but the USB keyboard control state and the sequence selection are not.
CAUTION To avoid the loss of data, GPIB settings, or current user instrument states that have not
been permanently saved to non-volatile memory, the MXG should always be powered
down either via the MXG's front panel power button or the appropriate SCPI command.
MXG's installed in rack systems and powered down with the system rack power switch
rather than the MXG's front panel switch display a Error –310 due to the MXG not being
powered down correctly.
Turning the USB keyboard control off disables the USB
keyboard; it has no effect on the Auto Recall softkeys.
For details on each key, use key help as described on see page 42.
USB Keyboard Key
Delete, Enter
Insert
Home, Up Arrow
End, Down Arrow
Page Up, Right Arrow, +
Page Down, Left Arrow, –
Signal Generator Function
RF On/Off
Mod On/Off
Next Sequence
Previous Sequence
Recall Next State Register
Recall Previous State Register