Agilent Technologies 62A Stereo System User Manual


 
140 Agilent N5161A/62A/81A/82A/83A MXG Signal Generators User’s Guide
Basic Digital Operation (Option 651/652/654) Preliminary
Waveform Sequences Preliminary
2. Change the first segment so that it repeats 100 times:
Highlight the first segment entry and press Edit Repetitions > 100 > Enter.
The cursor moves to the next entry.
3. Change the repetition for the selected entry to 200:
Press Edit Repetitions > 200 > Enter.
4. Save the changes made in the previous steps:
Press More > Name and Store > Enter.
To save the changes as a new sequence:
a. Press More > Name and Store > Clear Text.
b. Enter a file name (for example, SINE100+RMP200).
c. Press Enter.
The edited sequence saves as a new waveform sequence.
Playing a Sequence
If you have not created a waveform sequence, refer to “Creating a Sequence” on page 138.
NOTE To play a waveform segment individually or as part of a waveform sequence, the segment
must reside in BBG media. See also, “Loading a Waveform Segment into BBG Media” on
page 134.
1. Select a waveform sequence:
a. Press Mode > Dual ARB > Select Waveform.
b. Highlight a waveform sequence (for this example, SINE100+RMP200) from the Sequence On
column.
c. Press Select Waveform.
The display shows the currently selected waveform (for example, Selected Waveform:
SEQ:SINE100+RMP200).
2. Generate the waveform:
Press ARB Off On to On.
This plays the selected waveform sequence. During the waveform sequence generation, both the
I/Q and ARB annunciators turn on, and the waveform modulates the RF carrier.
Current waveform selection
Annunciators display with active waveform (ARB On)