6/08 ST 3000 Release 300 and SFC Model STS103 User’s Manual 85
6.8 Setting Range Values Using Local Adjustments,
Continued
Procedure, continued
Table 27 Setting Range Values Using Local Zero and Span
Adjustments, continued
Step Action
3
Loosen end-cap lock and remove end-cap from PWA side of
electronics housing to expose Local Zero and Span assembly or
Local Smart meter with Zero and Span adjustments.
Example – Local Zero and Span Assembly.
SPAN
ZERO
Example –Local Smart Meter with Zero and Span adjustments.
%
1000
UPPER
VALUE
UNITS
LOWER
VALUE
SET
VAR
SEL.
SPAN
ZERO
Continued on next page