9.5 Replacing Meter Body, Continued
Procedure, continued
Table 56 Replacing Meter Body Only, continued
Step Action
10
Feed flex tape on new meter body through neck of housing and
screw new meter body into housing until bottom of header portion of
center section is approximately flush with neck of electronics
housing.
11
Tighten outside set screw to be sure it is fully seated in slot in
header. Loosen set screw half turn, rotate housing to desired position
and tighten set screw.
12
Reverse actions in Steps 2 and 3 in Table 52, as applicable, to return
PWA and bracket to transmitter housing.
We recommend that you lubricate end-cap O-ring with silicon grease
such as Dow Corning #33 or equivalent before you replace end cap.
ATTENTION
Be sure to orient Local Smart Meter for proper
viewing through end-cap window. You can rotate the meter mounting
orientation in 90 degree increments.
13
Return transmitter to service and turn ON power.
14
Verify transmitter’s configuration data. Restore saved database, if
applicable.