Honeywell ST 3000 Satellite Radio User Manual


 
10 ST 3000 Release 300 and SFC Model STS103 User’s Manual 6/08
1.3 Smart Field Communicator, Continued
SFC model
differences, continued
Table 2 SFC Model Differences, continued
If SFC
model is. . .
Then it is compatible
with. . .
And additional functions
include . . .
STS102
Analog and Digital (DE)
mode ST 3000 pressure
transmitters, STT 3000
temperature transmitters,
and MagneW 3000
electromagnetic flowmeters.
Changing the mode from analog
to digital or digital to analog.
Configuration parameters for
Magnew 3000 as well as scratch
pad configuration area.
STS103
Same as STS102 plus new
multivariable transmitters -
SCM 3000 Smart Coriolis
Flowmeter and SGC 3000
Smart Gas Chromatograph.
SMV 3000 Smart
Multivariable Transmitters, if
SFC software version is 4.2
or greater.
SMV 3000 with superheated
steam algorithm and
thermocouple input, if SFC
software version is 4.4 or
greater.
Release 300 Series 100 and
900 ST 3000 pressure
transmitters, if SFC software
version is 5.0 or greater.
Two-line, 16-character per line
display. Made “
SAVE” and
RESTORE” functions part of
configuration menu instead of
dedicated keys. Configuration
parameters for SCM 3000 and
SGC 3000.
Configuration parameters for
SMV 3000
SMV 3000 configuration
parameters for superheated steam
algorithm and thermocouple
inputs.
Local Smart Meter configuration
parameters.
ATTENTION
The keystroke actions and prompt displays referenced in this manual are
for the SFC model STS103. While the SFC model STS103 does have a
two-line instead of a one-line display, many of the basic keystrokes and
configuration parameter prompts for ST 3000 pressure transmitters are
identical to those in the model STS102.
If you will be using a model STS102 SFC, you must refer to the SFC
Smart Field Communicator Operating Guide 34-ST-11-10 for keystroke
details. But, be aware that transmitter functions will be limited to only
those that are supported by the Model STS102 SFC.