Honeywell ST 3000 Satellite Radio User Manual


 
6/08 ST 3000 Release 300 and SFC Model STS103 User’s Manual 165
8.6 Monitoring Local Smart Meter Display, Continued
Typical operation
indications,
continued
Table 51 Summary of Typical Local Smart Meter Indications.
Meter Indication What It Means Meter Indication What It Means
%
100
0
No power
applied.
%
100
0
---
Meter has
detected
transmitter output
that is not-a-
number.
%
100
0
20
ANALOG
In H O
2
0
Normal display
for transmitter in
Analog mode
with digital
readout in
inches of water.
%
100
0
O
-
L
K
GPM
Display range is
Over Limit. Upper
value is
19,990,000 and
transmitter output
is over 100%.
%
1000
FLOW
GPM
K
99 90
Normal display
for transmitter in
DE mode and
square root
output. Digital
readout is
gallons per
minute with
1000 multiplier.
%
100
0
100
.
0
%
OUTPUT MODE
Transmitter is in
output mode.
Bargraph and
readout show
value that was
entered through
SFC.
%
100
0
77
.
9
%
CHECK STATUS
Transmitter in
DE mode is in
non-critical
status.
Displayed value
may not be valid.
If display is “- - -”
instead of a
value,
transmitter is in
critical status.
%
100
0
200
.
0
%
Input pressure
equal to or
greater than
200%. Display
flashes between
200% (or twice
current URV in
EU) and O-L.
Transmitter locks
output at 200%
and will go no
higher regardless
of input.