ST 3000 Release 300 and SFC Model STS103 User’s Manual 6/08
198
11.3 Clearing the “#” Symbol From SFC Display, Continued
Procedure, continued
Table 60 Clearing the # Symbol from the SFC Display, continued
If Message is. . . Then, Possible Cause is. . . And, Suggested Corrective
Action is. . .
#T 011P3
OAD.OM. E
STATUS
RVBL
OR
#T 011P3
AULROET Y
STATUS
DEFMB T
Pressure input is two times
greater than URL of transmitter.
• Check range and, if required,
replace transmitter with one that
has a wider range.
• Meter body may have been
damaged. Check the transmitter
for accuracy and linearity
Replace meter body and
recalibrate, if needed.
#
T 0 1 1 P 3
C O M A O E
S T A T U S
T D P N C P M
No temperature compensation
data exists for D/A converter.
Effect will be minor degradation of
ambient temperature influence
specifications. Replace electronics
module
#T 011P3
TEMOEV
STATUS
OSRSR PEN
Meter body temperature is too
high. Accuracy and life span may
decrease if it remains high.
Take steps to insulate meter body
from temperature source.
#T 011P3
ON
STATUS
NUWKN
Selection is unknown Be sure SFC software is latest
version. Press SHIFT and 3 to view
SFC software version.