Performance Tests
Performance Verification
4Ć26
b. Set function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 0ĂV
Offset 5.0ĂVĂ
(The actual voltage out is
1
/
2
of what is displayed.)
Output Off
2. Enable the function generator output: Turn on the function generator
output.
3. Check external ADD operation:
a. Check Autostep Step 4:
H Push the AWG2005 MANUAL TRIGGER button, and check that
the step changes to Step 4 on the MODE menu.
H Check that the DMM reading is in the range from 4.750 to
5.250ĂV.
4. Disable function generator output: Turn the function generator output off.
5. End procedure: Keep the test connections and instrument settings for
the next check.
Check Internal Add Operation
Electrical Characteristic Checked: Arithmetic Operation, Add, on
pageĂ1Ć13.
Equipment Required: Two 50ĂΩ coaxial cables, a function generator, and
a digital multimeter (DMM).
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1. Use test hookup and control settings from previous check.
2. Check internal ADD operation:
a. Check Autostep Step 5:
H Push the AWG2005 MANUAL TRIGGER button, and check that
the step changes to Step 5 on the MODE menu.
H Check that the DMM reading is in the range from 4.750 to
5.250ĂV.