Self Tests
Performance Verification
4Ć4
Figure 4Ć1:ăDiagnostics Menu
b. Run the diagnostics: Select Execute from the side menu. This
executes all the AWG2005 diagnostics automatically.
c. Wait: The internal diagnostics do an extensive verification of
AWG2005 functions. While this verification progresses, the screen
displays the clock icon. When finished, the resulting status appears
on the screen.
d. Confirm that no failures are found: Verify that no failures are found
and reported onĆscreen. If the diagnostics displays FAIL as the result
of any test, use the instructions in section 6, Troubleshooting, to
identify the faulty module. If the diagnostics display an error code,
contact your nearest representative.
2. Return to regular service: Push a button (other than UTILITY) in the
MENU column to exit the diagnostic menu.