Removal and Installation Procedures
Maintenance
6Ć22
EMI Gaskets
1. Remove front cover and trim ring: Do the Front Cover, Trim Ring, and
Menu Buttons procedure on page 6Ć22, removing only the front cover
and trim ring.
CAUTION
When reinstalling the EMI gaskets and/or the AWG2005 cabinet
carefully follow the instructions given to prevent exceeding the
environmental characteristics for EMI.
2. Remove EMI gaskets:
a. You will need needleĆnose pliers to do this part of the procedure.
b. Locate the EMI gaskets in the locator diagram, External Modules,in
Figure 6Ć3.
c. Use a pair of needleĆnose pliers to remove the four sections of EMI
gaskets from the groove in the trim ring.
3. Reinstall EMI gaskets: Press the EMI gaskets back into the groove in the
trim ring.
Front Cover, Trim Ring, and Menu Buttons
1. Assemble equipment and locate modules to be removed: No tools are
needed. Locate the modules to be removed in the locator diagram,
External Modules, in Figure 6Ć3.
2. Orient instrument: Set the AWG2005 with the back down on the work
surface and bottom facing you (see Figure 6Ć8).
3. Remove front cover: If the optional front cover is installed, grasp the front
cover by the left and right edges and snap it off of the trim ring. (When
reinstalling, align and snap back on.)
4. Remove frontĆpanel knob: Do the FrontĆpanel Knob procedure, on
pageĂ6Ć17.