Removal and Installation Procedures
Maintenance
6Ć18
Line Fuse and Line Cord
1. Assemble equipment and locate modules to be removed: You will need a
flatĆbladed screwdriver to do this procedure. Locate the line fuse and
line cord in the locator diagram, External Modules, Figure 6Ć3.
2. Orient instrument: Set the AWG2005 with the bottom down on the work
surface and the back facing you. If you are servicing the line fuse, do the
next step; if you are servicing the line cord, skip to step 4.
3. Remove line fuse: Find the fuse cap on the rear panel. See Figure 6Ć6.
Remove the fuse cap by turning it counterĆclockwise using a flatĆbladed
screwdriver, and remove the line fuse. Reverse the procedure to reinĆ
stall.
4. Remove line cord: Find the line cord on the rear cover. See Figure 6Ć6.
Remove the lineĆcord retaining clamp by first unplugging the line cord
from the line cord receptacle (1). Grasp both the line cord and the
retaining clamp and rotate it 90 degrees counterĆclockwise (2). Pull the
line cord and clamp away to complete the removal (3). Reverse the
procedure to reinstall.
5. Reinstallation: Do steps 3 and 4 in reverse order to reinstall the line cord,
and then the line fuse.