Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
AWG2005 Service Manual
4Ć37
Vertical CH1
CH1 Coupling DC
CH1 Scale 500ĂmV/div.
CH1 Input Impedance 50Ă
Horizontal
Sweep 50õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500ĂmV
Mode Auto
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Select waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to highlight the MRK.WFM file.
H Push ENTER to select the file.
3. Check rearĆpanel CH1 MARKER OUT amplitude:
a. Check CH1 MARKER OUT pulse amplitude:
H Check that the pulse amplitude of the displayed waveform is
2ĂV
pĆp
or greater.
4. Check rearĆpanel CH2 MARKER OUT pulse amplitude:
a. Check CH2 MARKER OUT pulse amplitude:
H Move the coaxial cable from the AWG2005 rearĆpanel CH1
MARKER OUT connector to the rearĆpanel CH2 MARKER OUT
connector.
H Check that the pulse amplitude of the displayed waveform is 2ĂV
or greater.
5. Check Option 02: If the AWG2005 has CH3 and CH4, repeat this entire
test, selecting the AWG2005 waveform and setting controls for CH3 and
CH4 and checking:
H RearĆpanel CH3 MARKER pulse amplitude
H RearĆpanel CH4 MARKER pulse amplitude
6. End procedure: Disconnect the oscilloscope.