Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
Performance Verification
4Ć40
OscilloscopeAWG2005
AWG2005
Function Generator
FG 5010
Figure 4Ć18:ăExternal Trigger Level Accuracy Initial Test Hookup
c. Set oscilloscope controls:
Vertical CH1
CH1 Coupling DC
CH1 Scale 0.2ĂV/div.
CH1 Input Impedance 50Ă
Horizontal
Sweep 50õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 0ĂV
Mode Auto
d. Set function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 0ĂV
Offset 0.6ĂV
Output Off
2. Select the AWG2005 waveform file and set AWG2005 controls:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Modify AWG2005 default settings:
H Push MODE!Gated!Polarity to highlight Positive.