Performance Tests
Performance Verification
4Ć16
6. End procedure: Turn off the function generator output, and disconnect
the function generator and oscilloscope.
Check Gated Mode
Electrical Characteristic Checked: Operating modes, Gated, on
pageĂ1Ć5.
Equipment Required: Three 50ĂΩ coaxial cables, a 50 Ω precision terĆ
mination, a function generator, and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2005 CH1 output
connector through the coaxial cable to the CH1 vertical input conĆ
nector on the oscilloscope.
b. Hook up the function generator: Connect the function generator
output to both the AWG2005 TRIGGER INPUT and the oscilloscope
CH2 input through a coaxial cable and a dual input coupler (see
Figure 4Ć5).
Function Generator
FG5010
OscilloscopeAWG2005
AWG2005
Figure 4Ć5:ăGated Mode Initial Test Hookup
c. Set oscilloscope controls:
Vertical CH1
CH1 coupling DC
CH1 scale 0.5ĂV/div.
CH1 input impedance 50ĂΩ