Performance Tests
AWG2005 Service Manual
4Ć13
b. Set the oscilloscope controls:
Vertical: CH1
CH1 coupling: DC
CH1 scale 0.2ĂV/div.
CH1 input impedance: 50ĂΩ
Horizontal
Sweep 50õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level -100ĂmV
Mode Auto
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Select the file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to display the list of waveform
files and highlight the MODE.WFM file.
H Push ENTER to select the file. This button is located to the
lowerĆright of the numeric keypad.
H Select the MODE.WFM file for CH2 same as CH1.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check against limits: Check that the amplitude of the sine wave disĆ
played on the oscilloscope is 5 vertical divisions and that 5 cycles of the
waveform are displayed.
5. Move the connection for the coaxial cable from the AWG2005 CH1
output to AWG2005 CH2 output connector.
6. Turn on the AWG2005 CH2 output: Push the CH2 button so that the LED
above the CH2 output connector is on.
7. Repeat procedure 2.
8. If Option 02 is installed (adds CH3 and CH4 output channels): Repeat
this procedure for CH3 and CH4 outputs.
9. End procedure: Disconnect the oscilloscope.