Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
AWG2005 Service Manual
4Ć15
Trigger
Source CH1
Coupling DC
Slope Positive
Level -100ĂmV
Mode Auto
d. Set the function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 4ĂV
Offset 2ĂV
Output Off
2. Set AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Modify the AWG2005 default settings:
H Push MODE!Triggered!Slope to select Positive slope.
H Select Level from the side menu and turn the general purpose
knob to select a 1ĂV trigger level.
c. Select the file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Highlight the MODE.WFM file using the general purpose knob.
H Push ENTER to select the file.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check triggered mode with manual triggering: Push the AWG2005
MANUAL TRIGGER button and check that when the button is pushed,
the oscilloscope displays a oneĆcycle sine wave.
5. Check triggered mode with external triggering:
a. Change the oscilloscope trigger mode to Normal.
b. Enable function generator output: Turn on the function generator
output.
c. Check triggering: Check that for each trigger supplied by the funcĆ
tion generator, the oscilloscope displays a oneĆcycle sine wave.