Performance Tests
Performance Verification
4Ć14
Check Triggered Mode
Electrical Characteristic Checked: Operating modes, Triggered, on
pageĂ1Ć5.
Equipment Required: Two 50ĂΩ coaxial cables, a function generator, and
an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2005 CH1 output
connector through the coaxial cable to the CH1 vertical input conĆ
nector on the oscilloscope.
b. Hook up the function generator:
H Connect the AWG2005 TRIGGER INPUT connector though a
coaxial cable to the function generator output connector (see
Figure 4Ć4).
OscilloscopeAWG2005
AWG2005
Function Generator
FG5010
Figure 4Ć4:ăTriggered Mode Initial Test Hookup
c. Set the oscilloscope controls:
Vertical: CH1
CH1 coupling: DC
CH1 scale 0.2ĂV/div.
CH1 input impedance: 50ĂΩ
Horizontal
Sweep 50õs/div.