Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
AWG2005 Service Manual
4Ć19
Procedure:
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2005 CH1 output
connector through the coaxial cable to the CH1 vertical input conĆ
nector on the oscilloscope (see Figure 4Ć7).
OscilloscopeAWG2005
AWG2005
Figure 4Ć7:ăWaveform Advance Mode Initial Test Hookup
b. Set oscilloscope controls:
Vertical CH1
CH1 coupling DC
CH1 0.2ĂV/div.
CH1 input impedance 50Ă
Horizontal
Sweep 50õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 0ĂV
Mode Auto
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Set AWG2005 controls:
H Push MODE!Waveform Advance!Slope to highlight PosiĆ
tive.
H Select Level from the side menu, and turn the general purpose
knob to select a 1.0ĂV level.
H Check that the side menu Run highlights Continuous. If necesĆ
sary, push Run to select Continuous.
c. Select waveform file: