Performance Tests
Performance Verification
4Ć42
6. End procedure: Turn off the function generator output and disconnect
the function generator.
This procedure checks the AWG2005 response to an external CLOCK IN
signal.
Electrical Characteristic Checked: Auxiliary Input, CLOCK, Threshold
level, on pageĂ1Ć15.
Equipment Required: Two 50Ăcoaxial cables, a function generator, and
an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Procedure:
1. Install test hookup and set test equipment controls:
a. Hook up oscilloscope: Connect the AWG2005 CH1 output through a
coaxial cable to the oscilloscope CH1 vertical input.
b. Hook up function generator: Connect the AWG2005 rearĆpanel
CLOCK IN through a coaxial cable to the function generator output
(see Figure 4Ć19).
AWG2005
Oscilloscope
Function Generator
FG5010
AWG2005
Figure 4Ć19:ăExternal CLOCK IN Initial Test Hookup
c. Set oscilloscope controls:
External CLOCK IN
Check