Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
AWG2005 Service Manual
4Ć17
Horizontal
Sweep 200õs/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500ĂmV
Mode Auto
d. Set function generator controls:
Function Square
Mode Continuous
Parameter
Frequency 1ĂkHz
Amplitude 4.0 V
Offset 2.0 V
Output Off
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Modify the AWG2005 default settings:
H Push MODE!Gated!Polarity to highlight Positive.
c. Select the file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Highlight the MODE.WFM file, using the general purpose knob.
H Push ENTER to select the file.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check gated mode with manual trigger: Push and hold the AWG2005
MANUAL TRIGGER button, and check that the oscilloscope continuĆ
ously displays a sine wave while the MANUAL TRIGGER button is
pushed.
5. Check gated mode with gate signal:
a. Change the oscilloscope controls:
Vertical CH1
CH1 coupling DC
CH1 scale 0.5ĂV/div.
CH1 input impedance 50Ă