Tektronix AWG2005 Stereo System User Manual


 
Performance Tests
Performance Verification
4Ć30
b. Set oscilloscope controls:
Vertical CH1
Coupling DC
Scale 500ĂmV/div.
Input impedance 50Ă
Horizontal
Sweep 500Ăns/div.
Trigger
Source CH1
Coupling DC
Slope Positive
Level 500ĂmV
Mode Auto
2. Set the AWG2005 controls and select the waveform file:
a. Initialize AWG2005 controls: Push UTILITY!Misc!Config...!ReĆ
set to Factory!O.K.
b. Select the waveform file:
H Push SETUP!Waveform Sequence, if necessary, to select a
waveform file for CH1. Waveform Sequence toggles between the
CH1 files (upper list) and the CH2 files (lower list).
H Turn the general purpose knob to select the CLK_AMPL.WFM
file.
H Push ENTER to select the file.
3. Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED
above the CH1 output connector is on.
4. Check clock amplitude accuracy: Check that the pulse amplitude of the
displayed waveform is 2 V
pĆp
or greater.
5. End procedure: Disconnect the oscilloscope.
This procedure checks the accuracy of the AWG2005 gain.
Electrical Characteristic Checked: Main Output, Amplitude, DC AccuĆ
racy, on pageĂ1Ć12.
Equipment Required: A 50Ă coaxial cable, a 50 termination, a BNCĆtoĆ
dual banana adapter, and a digital multimeter (DMM).
Prerequisites: The AWG2005 meets the prerequisites listed on page 4Ć7.
Gain Accuracy Check