AWG2005 Service Manual
1Ć11
Warranted Characteristics
This section will describe the warranted characteristics of the AWG2005.
These can be divided into two main categories: electrical characteristics and
environmental characteristics.
The electrical characteristics are valid under the following conditions:
1) The instrument must have been calibrated at an ambient temperature
between +20_ C to +30_ C.
2) The instrument must be in an environment whose limits are described in
Environmental Characteristics.
3) All tolerance limits apply after a 20 minute warm up and an execution of
the self calibration.
4) The instrument is operating at an ambient temperature between
+10_ C to +40_ C, unless otherwise noted.
Electrical Characteristics
Characteristics
Description Performance
Test
Clock Generator
Accuracy Check Clock
Frequency
A
+15_ C to +30_ C
0.005%
Accuracy,
page 4Ć27.
+10_ C to +40_ C 0.01%
Reference Oscillator
Accuracy "1ppm
Stability "1 ppm/year (20_ Cto30_ C)
Clock Generator (Option 05)
Accuracy Check Clock
Frequency
A
+15_ C to +30_ C
0.0005%
Accuracy,
page 4Ć27.
+10_ C to +40_ C 0.001%
Reference Oscillator
Accuracy "1ppm
Stability "1 ppm/year (20_ Cto30_ C)
Performance
Conditions