Characteristics
AWG2005 Service Manual
1Ć13
Electrical Characteristics
Characteristics Description Performance
Test
AM and Add
AM CH1 : Sine (512 points), Clock 20 MHz, AM, No Filter,
Amplitude 5 V, No Offset
CH2 : Waveform Data; FFF, No Filter, Amplitude 10 V
Internal Check InterĆ
nal AM Op
Amplitude Accuracy
Add 5% to the Amplitude Specification
*CH1 Out = CH1 x CH2,
nal AM Op
Ć
eration,
page 4Ć25.
Frequency Response DC to 10 MHz (-3 dB)
External Check ExterĆ
nal AM Op
Sensitivity 2 Vp-p (" 5%) signal causes 100% modulation.
*CH1 Out = CH1 x Ext Signal
Ext Signal 1V, 100% modulation
0V, 50% modulation
-1V, 0% modulation
nal AM Op
Ć
eration,
page 4Ć22.
Frequency Response DC to 10 MHz (-3 dB)
Add CH1 : sine (512 points), Clock 20 MHz, Add, No Filter,
Amplitude 5 V, No Offset
CH2 : Waveform Data; FFF, No Filter, Amplitude 2 V
Internal Check InterĆ
nal Add Op
Amplitude Accuracy
Add 5% to the Amplitude Specification
*CH1 Out = CH1 + CH2
nal Add Op
Ć
eration,
page 4Ć26.
Frequency Response DC to 10 MHz (-3 dB)
External Check ExterĆ
nal ADD Op
Amplitude Accuracy
Add 5% to the Amplitude Specification
*CH1 Out = CH1 + Ext Signal
nal ADD Op
Ć
eration,
page 4Ć25.
Frequency Response DC to 10 MHz (-3 dB)
*Note: Output Amplitude can't exceed 10 Vp-p (into 50W)